Spectrometer and spectroradiometer SR900

The SR900 is a compact spectroradiometer for spectrally resolved measurements in the UV, the visible range and the IR – on UV lamps, LEDs, irradiation systems and luminaires. Unlike a broadband radiometer it disperses the radiation and therefore makes lamp types with different spectra comparable, without a spectral mismatch error distorting the result. An array spectrometer without moving parts and a silicon detector with 2048 pixels deliver 2.3 nm resolution (FWHM) at 0.44 nm pixel spacing; the low stray-light design preserves sensitivity in the UV.

Reference spectra of common UV sources are collected in the UV spectra database.

The SR900 is combined with an SMA-905 quartz light guide and our radiometric sensor heads for measuring the irradiance and illuminance.

Traceable to the PTB, the SR900 is factory-calibrated. A DAkkS test in our accredited laboratory is optionally available. The SR900 thus enables exact spectroradiometric measurements for the assessment of irradiances and illuminances, biological effectiveness and color measurements. Due to the nearly scatter-free construction, a high sensitivity in the UV range is reached. Control and data evaluation is carried out by the spectral software SRpro. 

For the SR900 we additionally offer a wide range of accessories, which are combinable with customized extensions.

For luminous flux measurements and LED measurements, the SR900 can be combined with our integrating sphere.

Highlights of spectrometer SR900

  • Spectral range 200-1100 nm
  • Nearly scatter-free grid with 300 lines/mm
  • Spectral resolution 0.4 nm and 2,3 nm bandwidth
  • Order filtering
  • Spectral software for the color measurement, radiometric and effects-based measurements
  • Simple validation by means of the user concept
  • Wide range of accessories

Applications of UV-VIS spectrometer & spectroradiometer

  • Spectroradiometry
  • Spectroradiometry
  • Online process control, quality control
  • Testing of light sources (lamps, LEDs)
  • Color measurements and color control
  • Measurement of the global radiation
  • Inspection of optical materials and components, e.g. light conductors
  • Qualitative and quantitative analyses in chemistry, pharmacy and biology

In the Markets & Applications section, the SR900 appears wherever a broadband measurement no longer suffices: classifying solar simulators in photovoltaics, narrowband primaries in light and display measurement, weighted action spectra in photobiology, separating the UV and VIS channels under ICH Q1B in photostability testing, and 172 nm and 222 nm sources in semiconductor manufacturing.

Did you know...?

What does a spectrometer measure?


The Spectral Database Explorer shows typical emission spectra of UV LEDs and UV lamps, together with parameters such as UV proportions, illuminance and erythemal irradiance – exactly what the SR900 also delivers for your own light source.

Technical data spectrometer / spectroradiometer SR900

Spectral range 200 - 1100 nm
Spectral resolution 2,3 nm (FWHM)
Gratting 300 l/mm, holografic
Focal length 75 mm
Blaze wavelength 250 nm
Detector SI
Pixel size 14 x 200 µm
Number of pixel 2048
Pixel distancev0,44 nm  
Resolution 16 bit
Integrations time 0,1 ms to 60 s
Max. Dynamic range 10^4:1 with 3 measurements
Interface USB
Dimensions 177 x 125 x 45 mm
Weight 0.6 kg
Supply voltage 5 V, 500 mA  
Power < 3 W
Operating temperature 5 to 40 °C
Storage temperature -10 to 60 °C
Humidity <80%, non-condensating
System requirements Windows 10/11, min 1 GB RAM